| Signal generating unit response time |
Approximately 1 sec. (Time between start of voltage change and when the voltage enters accuracy range) |
| Signal genertating unit voltage limiter |
Approximately 32 Volts |
| Signal genertating unit current limiter |
Approximately 25 mA |
| Divided output (m/n) function |
output= setting x (n/m) x n=0-m; m=1-9; n ≤ m |
| Auto-step output function |
n value sent automatically when n/m value selection is slected (two option: approximately 2.5 seconds/step or 5 seconds/step ) |
| Sweep function |
Sweep time (two option: approximately 16 seconds or 32 seconds) |
| Memory function |
50 value sets (generated and measured values are storedas value set with the same address (up to 50 value sets can be strored)) |
| Measuring unit maximum input |
Voltage terminal: 300VAV AC Current terminal: 120 mADC |
| Current terminal input protection |
fuses 125 mA/250V |
| Measuring unit ground voltage |
300 VAC maximum |
| Measurement display update rate |
Approximately once per second |
| Display |
Segmented LCD (approximately 76mm x 48 mm ) |
| Backlight |
LDE Backlight; auto-off after 1 minute ( from when the light key is turned on) |
| Consumed power |
Approximately 7VA ( using 120Vac Adapter) |
| Auto-power off function |
Approximately 10 minutes |
| Applicable standards |
IEC6101-I; IEC6101-2-31 EM61326: 1997 +1A: 1988 EN55011: 1988, Class B, Group 1 |
| Insulation resistance |
Across input and output terminal, 500 VDC, 50MΩ or greater |